東京プラント株式会社 Sitemap EnglishJapanese

Products

Combustion analysis system OSIRIS

Hardware specification
Model OSIRIS Evolution3
PC communication Dedicated Ethernet gigabit
Analog Input Input: 8 ch (ADC 18 bits, 1 MHz)
Time sampling: up to 500kHz
Voltage range: ±10 V, ±5 V, ±2 V, ±1 V
External voltage input: 2 ch ±50 V, ±25 V
External input: 2 ch IEPE (e.g.: accelerometer), or AC (e.g.: microphone)
Digital channel Digital output: 4 ch relay, TTL
Digital input: 4 ch HTTL
Synchronization Encoder input: 1LVDS TTL, RS422
Sensor input: 2 ch HTTL, analog
Synchronization output: 3 ch (analog, frequency, speed)
Target management
(FFR-M)
Up to 4 singularities per 1 rotation, different markings (short/long)
Incl. additional markings. Output frequency: up to 1.5 MHz
Power supply 9-30 VDC (6-48 VDC supported for transient voltage)
Power consumption 14W
Dimension (mm) 220 x 42 x 140 (1U, 9.5 inch)
CEM ICE61326-1
Operating temperature -40 to +50 °C
Extension 4 x Evolution3 modules (32 ch)
up to 8 ACPM charge amplifiers
Software specification
Measurement mode Angle: simple, at event (before/after event), cool start, angle and time
Time (1 to 500kHz)
TDC offset Cylinder pressure or set from TDC signal
File format XODF, IFILE, ASCII (incl. custom and window function)
Graphical display Engine cycle, P/V, X/Y, tendency, monitoring, distribution, spectrum
Monitoring Monitoring by fixed or statistic values at each calculated value.
Alarm feature Monitoring the upper/lower limit of each signal and calculated item.
Communication interface AK via DCOM interface, INDI, TCPIP, RS232
Charge amplifier (option) ACPM, D2T, SCP Kistler, MICRO IFEM AVL
Other I/O (option) CAN bus I/O, ASAM-MCD3 input, analog output
Cylinder pressure calculation PMAX, IMEP, dP/dΘ, engine speed…
Statistic value calculation min., max., average, standard deviation, COV…
Combustion analysis Heat release, burned fraction (CA10, CA50…), polytropic coefficient, thermal losses
Knock (option) Calculation based on energy (FFT), pressure variation (MAPO, KP)
Noise (option) Noise level, third octave structural attenuation
Other (option) Cycle variation, FFT, needle lift, FIR filtering, low/high and band-pass, injector timing (current clamp)
Global calculation Multicylinder, script calculation by VB…
Custom calculation C++ (Training from FEV enginner to be necessary)

*Specification is subject to change due to product modification.